I would appreciate your guidance regarding a medication safety and system limitation issue encountered in our NICU. We received an order for:
• Insulin Regular 100 units/mL injection.
• Prescribed dose: 0.025 units IV (one-time dose)
• Context: Medication supplied from floor stock, with nursing responsible for preparation and administration.
When attempting to verify the order in Cerner, the system generated the following error: Calculated dose: 0.025 units / 0.00025 mL “A strength less than 0.0001 or a volume less than 0.01 was entered or calculated. Enter a larger value or modify the units.”
As the calculated administration volume (0.00025 mL) is significantly below the system’s minimum threshold, the order cannot be processed through the pharmacy verification workflow.